Magnet ID :     Calibration
Directory : S:\magdata\LCLS-II\Phase Shifter\Calibration\126Gcm_+By_downstream\Final Results\Stretched Wire
   
Date : 06-29-2018
Time : 12:34:58 - 13:39:19
Operator : VV
   
Project : LCLS-II
Device Type : Phase Shifter
Run Number : 1
Data Type :
Run Comment : 126 Gauss*cm magnet, +By, downstream pos
   
gap I1Y I2Y I1X I2X
-
110.3 0.4 (4)
6.1 0.3 (5)
4.9 0.5 (5)
1.2 0.4 (5)

integral gap -02.00;+00.00 -01.00;+00.00 +00.00;+00.00 +01.00;+00.00 +02.00;+00.00
I1Y -
108.1 0.6 (3)
109.2 0.8 (3)
108.3 0.6 (3)
111.1 2.7 (3)
108.8 0.4 (3)
I2Y -
5.9 0.3 (3)
6.1 0.4 (3)
5.9 0.3 (3)
4.9 2.3 (3)
6.1 0.5 (3)
I1X -
4.9 0.6 (3)
5.2 0.5 (3)
4.9 0.5 (3)
5.0 0.5 (3)
5.1 0.6 (3)
I2X -
0.4 0.0 (3)
0.7 0.3 (3)
0.4 0.0 (3)
0.9 0.2 (3)
0.6 0.4 (3)

integral gap +00.00;-02.00 +00.00;-01.00 +00.00;+00.00 +00.00;+01.00 +00.00;+02.00
I1Y -
109.7 0.5 (3)
109.7 0.5 (3)
109.6 0.6 (3)
110.0 0.3 (3)
109.9 0.6 (3)
I2Y -
6.3 0.2 (3)
6.3 0.2 (3)
6.4 0.1 (3)
6.3 0.2 (3)
6.4 0.1 (3)
I1X -
4.7 0.3 (3)
1.3 0.9 (3)
3.3 2.7 (3)
2.3 1.0 (3)
2.3 2.4 (3)
I2X -
0.6 0.2 (3)
0.2 0.3 (3)
0.5 0.5 (3)
0.3 0.1 (3)
0.3 0.4 (3)

   
Magnet ID :     Calibration
Directory : S:\magdata\LCLS-II\Phase Shifter\Calibration\126Gcm_+By_downstream\Final Results\Stretched Wire